3.1 Setup Reminders
3.1.1 Average Installation Time
Once the site has been prepared, the average installation time required is shown in Table 3-1 below.
Table 3-1 average installation time
Description Average Installation Time Comments
Unpacking the system 0.5 hours
Installing the system / options / printers 0.5 to 1.5 hours depends on required configuration
DICOM Option (connectivity) 0.5 to 1.5 hours depends on configuration amount
Install InSite 0.5 hours
3.1.2 Installation Warnings
Caution
Since the Voluson E-Series weighs approximately 150 kg (330 lbs.) without peripherals, two people are
required to unpack it.
Warning
There are no operator serviceable components. To prevent shock, do not remove any covers or panels.
Should problems or malfunctions occur, unplug the power cord.
Only qualified service personnel should carry out servicing and troubleshooting.
3.1.2.1 Moving/Lifting the System
Note
For important safety considerations see Section 1.5.3 "Mechanical Safety" on page 1-20 .
How to lift the system:
1 Preparation
Disconnect all probes and transport them separately.
Disconnect the ECG cable (if applicable) and transport it separately.
Ensure all peripheral devices (printer, ...) are firmly fixed within the system.
2 Remove the footrest/wheel axis cover on the front side of the system: Turn the 4 quick release devices below the footrest 90°
(see:
Figure 8-21 on page 8-17
).
3 Pass a strap through the openings in the metal sheet.
4 Lift the system by the straps and the rear handle.
Caution
Do not pull or lift the system with the front handle of the user interface (control console).
Always use a strap to lift the system. Do not grasp the metal sheet with your hands.
Move the system forward or backward when going up or down inclines. Do not move the system
sideways or diagonally.
Setup Instructions
3-2
Voluson E-Series Service Manual
5539550APB Revision 6

Table of Contents