8.2 System Software - Installation/Upgrade procedure
Introduction
There are two possibilities to update the system software:
1.
via the
FMI from DVD
button in the System Setup
Service
page
-
Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS patches, etc.) and installation time depend on contents of the used System DVD.
-
If the currently installed software has to be upgraded to a newer version, a new software specific
"Permanent key" is required. Please contact your local distributor or GE service representative.
2. by downloading and installing a
Software Update Package
via active InSite connection
-
Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS patches, etc.) and installation time depend on contents of the SW update package.
-
PRECONDITION: InSite connection has to be configured and checked out. Operation see:
Section 3.13.8 "InSite Configuration" on page 3-70
.
Caution
Disconnecting ALL external USB devices (except DVD drive) is NECESSARY. Re-installation of any
previously attached printer has to be done after the upgrade procedure.
Note
Installing the Bluetooth Printer and its connection set is NOT possible by the user.
Manpower
One person ~ 1 hour (depends on contents of System DVD resp. SW update package, peripherals, etc.)
Tools
System DVD
Replacement Procedures
Voluson E-Series Service Manual
5539550APB Revision 6
8-3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E Series and is the answer not in the manual?

GE Voluson E Series Specifications

General IconGeneral
BrandGE
ModelVoluson E Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals