Item Service at indicated Time D W M A Notes
Measurement Accuracy
Checks
Also after corrective maintenance or
as required by your facilities QA
program.
Probe/Phantom Checks Check gain changes, vary the focus and check
reaction on screen.
NOTE! The use of a Phantom is not required during
Preventive Maintenance. Customer may use it as part
of their Quality Assurance Program tests.
Also after corrective maintenance or
as required by your facilities QA
program.
Functional Checks of all
probes
Section 10.4.2 on
page 10-5
Also after corrective maintenance or
as required by your facilities QA
program.
10.3 Tools required
10.3.1 Special Tools, Supplies and Equipment used for Maintenance
Note
Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant
with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551 or IEC 62353 or other relevant national regulation.
Table 10-2 overview of supplies and equipment used for maintenance
Tools Comments
Digital Volt Meter (DVM) minimum 5% accuracy, 3.5 digit and 200 Ohm range required
Anti Static Kit including anti-static mat, wrist strap, cables and conductive ground cord
Anti Static Vacuum Cleaner
QIQ Phantom
NOTE! The use of a Phantom is not required during Preventive Maintenance. Customer
may use it as part of their Quality Assurance Program tests.
CD-RW Media (minimum quad speed)
DVD+RW Disc Media blank blank 4,7GB DVD+RW disc
B/W Printer Cleaning Sheet see printer user manual for requirements
Color Printer Cleaning Sheet see printer user manual for requirements
Disposable Cloves
Care and Maintenance
10-4
Voluson E-Series Service Manual
5539550APB Revision 6