10.6.4 Grounding Continuity
Danger : Electric Shock Hazard!
The patient or operator MUST NOT come into contact with the equipment during this test.
Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case. The
ground wire resistance should be less than 0.2 ohms. Reference the procedure in the IEC60601-1-1 and/or
IEC62353.
Figure 10-2 Ground Continuity Test
1 Ultrasound System 3 Ohmmeter
2 Ground Pin 4 accessible Metal Parts such as:
Potential equilibrium connector
Monitor housing
Probe connector
Caution
Lacquer is an isolation barrier! Measure only on blank accessible metal parts.