6.3.2 Height Adjustment (Elevation)
Height adjustment is done with the Lift UP / Lift DOWN button inside of the handlebar opening.
As long as a button is pressed, the control console can be lifted / lowered.
Caution
Make sure that nothing would be jammed while moving!
6.3.2.1 Moving the Console without booting up the System
If it is impossible to boot up the system, the user interface can be lowered and/or lifted by pressing 3 keys on
the control console.
1.
Connect the main power cable to the back of the system.
2. Connect the main power cable to a hospital grade power outlet with the proper rated voltage.
3. Press 3 keys (see
Figure 6-4 below
) on the control console simultaneously to move it
-
downwards: Patient ID key, End Exam key and Lift DOWN button
-
upwards: Patient ID key, End Exam key and Lift UP button
Figure 6-4 keys to lower/lift the control console
Note
Shipping the Voluson E-Series ultrasound system in its original packaging is only possible when control
console is centered and locked in position, the system is lowered to its minimum height and the monitor is
flapped down and locked (see Figure 6-1 on page 6-3 ).
6.4 Modification of Keyboard Layout
Modification of the Keyboard Layout is no more possible at systems with operating system Windows 10
(software version EC330, BT18).
Note
Configuring the layout of the keyboard is only useful by changing the concerned keys also; see Section 8.13
"Replacement of Key Caps (by special native language keys)" on page 8-24 .
1. Press the Utilities key on the control console.
2. In the “Utilities” menu touch the
Setup
button to invoke the setup desktop on the screen.
3. On the right side of the screen select
Administration
and then click the
Service
tab.
4. Enter the password « she » and click the
Accept
button to display the Service Tools page (see
Figure
5-27 on page 5-50
).
Service Adjustments
Voluson E-Series Service Manual
5539550APB Revision 6
6-5