8.16 Replacement of Fuses at Power Supply Module (RSP)
Manpower
One person, 30 minutes
Tools
small-sized slotted screwdriver
Preparations
1.
Power Off/Shutdown the system as described in
Section 4.2.2 on page 4-4
.
8.16.1 Fuses at Power Supply Module (RSP) - Replacement Procedure
1. Open the fuse protection at the power inlet block (1) with a small-sized slotted screwdriver.
Figure 8-33 exchange fuses
2. Remove the fuse holder and exchange fuses.
Replacement Procedures
Voluson E-Series Service Manual
5539550APB Revision 6
8-27