8.8 Replacement of Covers
Content in this section
8.8.1 Replacement of the Footrest Cover & Wheel Axis Cover - - - - - - - - - - - - - - - - - - - 8-17
8.8.2 Replacement of the Front Frame - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-18
8.8.3 Replacement of the Air Filter Cover - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 8-19
8.8.1 Replacement of the Footrest Cover & Wheel Axis Cover
Manpower
One person, 1 minute
Tools
none
Preparations
1. Power Off/Shutdown the system as described in
Section 4.2.2 on page 4-4
.
8.8.1.1 Footrest Cover & Wheel Axis Cover - Removal Procedure
1. For unlocking, turn the 4 quick release devices below the footrest 90°.
Figure 8-21 turn quick release devices 90° and remove the footrest cover
2. Remove the Footrest Cover & Wheel Axis Cover.
8.8.1.2 Footrest Cover & Wheel Axis Cover - Installation Procedure
1. Place the (new) Footrest Cover & Wheel Axis Cover on the original position.
2. For locking, turn the 4 quick release devices below the footrest 90°.
Replacement Procedures
Voluson E-Series Service Manual
5539550APB Revision 6
8-17

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E Series and is the answer not in the manual?

GE Voluson E Series Specifications

General IconGeneral
BrandGE
ModelVoluson E Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals