10.6.2 Leakage Current Limits
The following acceptance limits and test conditions are summarized from IEC 60601 including national
deviations and IEC 62353 and in some cases are lower than that specified by the standards.
Table 10-10 leakage current limits
Leakage Current Test System Power Grounding/PE Conductor Limit mA Limit μA
Earth Leakage On and Off N/A 0.5 500
Chassis/Enclosure Leakage On and Off Closed
Open
0.1
0.5
100
500
Type BF Applied Parts On (transmit) Closed
Open
0.1
0.5
100
500
Type CF Applied Parts On (transmit) Closed
Open
0.01
0.05
10
50
Type BF Applied Parts (sink leakage,
mains voltage on applied part)
On and Off Closed 5.0 5000
Type CF Applied Parts (sink leakage,
mains voltage on applied part)
On and Off Closed 0.05 50
Values based on IEC 60601
Note
Open Grounding is also known as “Lift Ground”.
Table 10-11 ISO and Mains Applied limits*
Probe Type Measurement
BF 5.0 mA (5000 μA)
CF 0.05 mA (50 μA)
* ISO and Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the
applied part to determine the amount of current that will pass (or sink) to ground if a patient is in contact with
mains voltage.
Note
Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant
with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551 or IEC 62353 or other relevant national regulation.