7.4.2.2 Export Log's and System Data (via Service Page)
1.
Press the Utilities key on the control console.
2. In the “Utilities” menu touch the
Setup
button to invoke the setup desktop on the screen.
3. On the right side of the screen select
Administration
and then click the
Service
tab.
4. Enter the password « she » and click the
Accept
button to display the Service Tools page (see
Figure
5-27 on page 5-50
).
5. Click on the
Export System Data
button to Full Backup the System state. This includes dump files and
text files, full Service Database informations about probes, boards, Software, Options and the Event
Log File.
7.4.2.2.1 Dump file
Every time an error message is produced, a dump file and a text file containing the error dump and the error
message are created in D:\export. Up to 20 dump files are stored there.
Figure 7-11 system has encountered a problem
After clicking on
OK
the system reboots automatically.