7.8.6.2 Record Spectral Doppler display
Note
The option SW DVR (Software Digital Video Recording) is required to perform this test.
1. Insert an empty, formatted DVD/CD+R/RW disc into the drive.
Note
If you use an unformatted DVD+RW, recording is impossible! To format the DVD, press the DVD key and
then select the Format button.
2. Press the PW key on the control panel to start PW preparation mode.
3. Activate Doppler motion display by pressing the right/left trackball key.
4. Press the DVD key on the control console once to display the remote control menu.
Figure 7-33 Remote Control menu
5. In the remote menu touch the
Record
button to record a Doppler motion display (e.g., of the
Carotid
).
6. Touch the
Stop
button to stop recording.
7. Touch the
Finalize
button. The recorded motion sequence is stored on DVD.
8. Touch the
Play
button to play-back the recorded Doppler motion sequence.
With the
Skip
buttons the next/previous sequence can be selected.
Limits: The recorded Doppler motion sequence is running. The audio signal of the arterial flow can be heard in the
loudspeaker(s).
Diagnostics/Troubleshooting
7-30
Voluson E-Series Service Manual
5539550APB Revision 6