8. After clicking
Install
, the SW update procedure starts.
The system saves Full Backup in silent mode on R:, then it reboots into LINUX.
Figure 8-13 installation in progress
9. Check disk is performed automatically - restart.
Figure 8-14 Check disk is performed automatically
A silent “Rollback” image from C:\ is stored on R:\. After executing all LINUX commands, the system
reboots again.
10.
3 dots (one after the other) appear on the screen.
11. Booting auto ....
Figure 8-15 Boot screen - auto