17-19 Operators’ Manual
990400 Rev C 2010 May 1
11-2
The Calibration screen is displayed.
3. Click the Calibrate button (top) in the Panel field.
WARNING
The X-ray device may be dangerous to the Patient and operator if
you do not observe and follow the operating instructions. Do not
operate this system unless you have received training to perform a
procedure.
A window is displayed “Remove all objects from the field of
view and click OK to start X-ray exposure.”
4. Click OK.
5. When prompted, press the Scan button on the Control Box. An
audible alarm is sounded and the X-ray ON light is illuminated
during radiation exposure.
6. The progress of the scan is displayed on the Progress bar at the
bottom of the Acquire window. You will be prompted to press
the Scan button several times as the calibration progresses.
“Panel Calibration Complete” is displayed when the process is
finished. Click OK.