EasyManua.ls Logo

Imaging Sciences i-Cat 17-19 - Power Requirements

Imaging Sciences i-Cat 17-19
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17-19 Operators’ Manual
990400 Rev C 2010 May 1
14-2
Image Detector: Amorphous Silicon Flat Panel (readable area), 9.37” width x
7.56” height (23.8 cm width x 19.2 cm height)
Sensor Front Panel Attenuation Value: Less than 1mm of aluminum
equivalent (information for reference only)
Gray Scale: 14 bit
Voxel Size: 0.4/0.3/0.25/0.2 mm
Image Acquisition: Single 360 degree rotation (maximum)
Scan Time: 26.9/8.9 seconds
Field of View: (standard) 6.50 inches diameter x 5.33 inches height
(maximum) (16.5 cm diameter x 13.5 cm height (maximum))
Extended Field of View: (optional) 9.15 inches diameter x 6.67 inches height
(maximum) (23.2 cm diameter x 17 cm height (maximum))
Note: Maximum values can be collimated down.
Primary Reconstruction: Less than 2 minutes for 26.9 second scan @
0.4 voxel
Secondary Reconstruction: Real Time
Stopping Distance and Angle: Hard stop is at -45
o
and 470
o
(reference is the
gantry at the home position being 0
o
). Platform travel is 69 mm.
Power Requirements
The Scanner requires a Dedicated Line. A Surge Protector is recommended.
The Scanner is suitable for continuous connection to a power supply in stand-
by mode.
Line Voltage: 100VAC, 115VAC, 200VAC or 230VAC (Factory Set)
Line Voltage Regulation requirement: + 10%
Line Current: 15 Amps (100V), 10 Amps (115V), 7.5 Amps (200V) or 5
Amps (230V)
Line Frequency: 50 Hz / 60 Hz
Phase: Single
Main Circuit Breaker: 15 Amps (100V), 10 Amps (115V), 7.5 Amps (200V),
or 5 Amps (230V)
Nominal Electrical Input Power to Supply: Volume Scan = 300W (120kV,
5mA); PAN Scan (Large) = 625W (94kV, 5mA). Scan Time has no affect on
electrical power output.
Nominal Maximum Electrical Input Power at 100 kV: 850W (100 kV,
7mA). Scan Time has no affect on electrical power output.