990400 Rev C 2010 May 1
11-1
Chapter
11
Calibration
The Panel, Collimators, and Geometry Calibrations can be
conducted by the Owner / Operator of the device. It is recommended
that the Panel Calibration be performed once a week.
Panel Calibration
The Panel Calibration is performed for both Portrait and Landscape
positions. The entire calibration takes approximately 8 to 10 minutes
to complete. The process performs a five mode calibration as listed
below:
Mode 0 (4 x 4) Landscape
Mode 1 (2 x 2) Landscape
Mode 2 (2 x 2) PAN (Landscape)
Mode 3 (4 x 4) Portrait
Mode 4 (2 x 2) Portrait
Mode 5 (1 x 1) PAN (Landscape, not supported)
To run the Panel Calibration:
1. Ensure room temperature is in the range of 50 to 95
o
F (10 to
35
o
C).
2. From the desktop, double click the Calibration
icon.