Product Information
14-5
Extension Cords
Do not use any extension cords which are not provided with the
system. Be aware that multiple portable socket outlets or extension
cords are not to be connected to the system.
External Item
Do not connect any items or equipment to this system which are not
part of the system.
Cleaning
Routinely clean and disinfect all items which come in contact with
the patient. Use Opti-Cide
3®
Solution and/or Wipes from Biotrol
International, or equivalent cleaner and disinfectant. Opti-Cide
3®
is
a broad-spectrum disinfectant that is effective in three minutes. See
Opti-Cide
3®
label for full instructions.
Electromagnetic or other Interference
(Emissions and Immunity)
The system was tested and it was determined that a ferrite bead was
necessary on the Ethernet cable in order to meet the class A (non-
residential) limits. With the system configured with the ferrite bead,
it complied with the criteria contained in IEC 60601-1-2 Edition 2.1
Issued 2004/11/01 and JIST0601-1-2 Issued 2002/07/025.