17-19 Operators’ Manual
990400 Rev C 2010 May 1
14-6
1
During the 5 second voltage dropout, the system powered down
and then, through normal operation of the controls, was able to be
restored to normal operation.
Test Name Test Level/
Equipment Class
Results /
Notes
Immunity
Performance
Criteria Met
Emissions Testing
Radiated Emissions Class A; Group 1 Compliant -
Conducted Voltage Emissions Class A; Group 1 Compliant -
IEC61000-3-2 Harmonic
Current Emissions
Class A Compliant -
IEC61000-3-3 Voltage
Changes, Voltage
Fluctuations and Flicker
Class A Compliant -
Immunity Testing
61000-4-2 Electrostatic
Discharge
+ 6 kV Contact, +8
kV Air
Compliant A
61000-4-3 Radiated Immunity 80 MHz - 2.5 GHZ,
3 V/M, 80% AM
with 1kHz
Compliant A
61000-4-4 Electrical Fast
Transients
+2 kV Power
Supply Lines, +
1
kV Input/Output
Lines
Compliant A
61000-4-5 Surge Immunity +1 kV Line to Line,
+
2kV Line to
Earth
Compliant A
61000-4-6 Conducted
Immunity
150 kHz - 80 MHz,
3 Vrms
Compliant A
61000-4-8 Power Frequency
Magnetic Field
3 A/M Compliant A
61000-4-11 Voltage Dips and
Short Interruptions
>95% dip for 0.5
periods
Compliant A
60% dip for 5
periods
Compliant A
30% dip for 25
periods
Compliant A
>95% dip for 5
seconds
Compliant
C
1