990400 Rev C 2010 May 1
Radiation Environment Survey
13-15
Operator Data
Available focal spot size (mm):
0.5
Available target angle: 15°
Anode construction:
Vacuum cast copper with
tungsten target
Cathode construction:
Vacuum tube nickel with
tungsten filament
Max. Tube operating voltage:
130 kVp Full wave rectified
Inherent filtration/Window
material:
2.0 mm/Glass
Cooling Method:
High dielectric transformer oil
Conditioning of Tube Head:
Must be at or above 22
o
C before
first scan.
No system parts are connected
to a Protected Earth Terminal.
X-Ray Tube Voltage
X-Ray Tube Current Normal Electrical Power
84 kV
5mA 420W
94 kV
5mA 470W
120 kV
5mA 600W
SXR-130-15-0.5
FOCAL SPOT
TO RECEPTOR

Table of Contents