108 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
5 Common (*) Commands
*OPC (see page 118) *OPC? (see page 118) ASCII "1" is placed in the output
queue when all pending device
operations have completed.
n/a *OPT? (see page 119) <return_value> ::= 0,0,<license
info>
<license info> ::= <All field>,
<reserved>, <reserved>,
<reserved>, <Memory>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<Segmented Memory>, <Mask Test>,
<reserved>, <Bandwidth>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>, <Educator's Kit>,
<Waveform Generator>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<reserved>, <reserved>,
<Digital Voltmeter>, <reserved>,
<reserved>, <reserved>, <Remote
Command Logging>, <reserved>,
<reserved>, <reserved>
<All field> ::= {0 | All}
<reserved> ::= 0
<Memory> ::= {0 | MEMUP}
<Low Speed Serial> ::= {0 | EMBD}
<Automotive Serial> ::= {0 |
AUTO}
<Segmented Memory> ::= {0 | SGM}
<Mask Test> ::= {0 | MASK}
<Bandwidth> ::= {0 | BW10 | BW20}
<Educator's Kit> ::= {0 | EDK}
<Waveform Generator> ::= {0 |
WAVEGEN}
<Digital Voltmeter> ::= {0 | DVM}
<Remote Command Logging> ::= {0 |
RML}
*RCL <value> (see
page 120)
n/a <value> ::= {0 | 1 | 2 | 3 | 4 |
5 | 6 | 7 | 8 | 9}
Table 39 Common (*) Commands Summary (continued)
Command Query Options and Query Returns