386 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:VAMPlitude
(see page 798)
Command Syntax
:MEASure:VAMPlitude [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VAMPlitude command installs a screen measurement and starts a
vertical amplitude measurement. If the optional source parameter is specified, the
current source is modified.
Query Syntax
:MEASure:VAMPlitude? [<source>]
The :MEASure:VAMPlitude? query measures and returns the vertical amplitude of
the waveform. To determine the amplitude, the instrument measures Vtop and
Vbase, then calculates the amplitude as follows:
vertical amplitude = Vtop - Vbase
Return Format
<value><NL>
<value> ::= the amplitude of the selected waveform in NR3 format
See Also • "Introduction to :MEASure Commands" on page 350
• ":MEASure:SOURce" on page 380
• ":MEASure:VBASe" on page 388
• ":MEASure:VTOP" on page 394
• ":MEASure:VPP" on page 391

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals