:DVM Commands 13
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 243
:DVM:CURRent
(see page 798)
Query Syntax
:DVM:CURRent?
The :DVM:CURRent? query returns the displayed 3-digit DVM value based on the
current mode.
Return Format
<dvm_value><NL>
<dvm_value> ::= floating-point number in NR3 format
See Also • ":DVM:SOURce" on page 247
• ":DVM:ENABle" on page 244
• ":DVM:MODE" on page 246
• ":DVM:FREQuency" on page 245
NOTE
It can take up to a few seconds after DVM analysis is enabled before this query starts to
produce good results, that is, results other than +9.9E+37. To wait for good values after DVM
analysis is enabled, programs should loop until a value less than +9.9E+37 is returned.