484 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:IIC:ASIZe
(see page 798)
Command Syntax
:SBUS<n>:IIC:ASIZe <size>
<size> ::= {BIT7 | BIT8}
The :SBUS<n>:IIC:ASIZe command determines whether the Read/Write bit is
included as the LSB in the display of the IIC address field of the decode bus.
Query Syntax
:SBUS<n>:IIC:ASIZe?
The :SBUS<n>:IIC:ASIZe? query returns the current IIC address width setting.
Return Format
<mode><NL>
<mode> ::= {BIT7 | BIT8}
Errors • "-241, Hardware missing" on page 759
See Also • "Introduction to :SBUS<n> Commands" on page 461