640 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
28 :WAVeform Commands
:WAVeform:BYTeorder
(see page 798)
Command Syntax
:WAVeform:BYTeorder <value>
<value> ::= {LSBFirst | MSBFirst}
The :WAVeform:BYTeorder command sets the output sequence of the WORD data.
• MSBFirst — sets the most significant byte to be transmitted first.
• LSBFirst — sets the least significant byte to be transmitted first.
This command affects the transmitting sequence only when :WAVeform:FORMat
WORD is selected.
The default setting is MSBFirst.
Query Syntax
:WAVeform:BYTeorder?
The :WAVeform:BYTeorder query returns the current output sequence.
Return Format
<value><NL>
<value> ::= {LSBF | MSBF}
See Also • "Introduction to :WAVeform Commands" on page 635
• ":WAVeform:DATA" on page 642
• ":WAVeform:FORMat" on page 644
• ":WAVeform:PREamble" on page 649
Example Code • "Example Code" on page 654
• "Example Code" on page 650