:RECall Commands 22
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 435
:RECall:FILename
(see page 798)
Command Syntax
:RECall:FILename <base_name>
<base_name> ::= quoted ASCII string
The :RECall:FILename command specifies the source for any RECall operations.
Query Syntax
:RECall:FILename?
The :RECall:FILename? query returns the current RECall filename.
Return Format
<base_name><NL>
<base_name> ::= quoted ASCII string
See Also • "Introduction to :RECall Commands" on page 433
• ":RECall:SETup[:STARt]" on page 438
• ":SAVE:FILename" on page 444
NOTE
This command specifies a file's base name only, without path information or an extension.