EasyManuals Logo

Keysight Technologies InfiniiVision DSOX1204A User Manual

Keysight Technologies InfiniiVision DSOX1204A
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #261 background imageLoading...
Page #261 background image
:FFT Commands 15
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 261
:FFT:CENTer
(see page 798)
Command Syntax
:FFT:CENTer <frequency>
<frequency> ::= the current center frequency in NR3 format. The range
of legal values is from -25 GHz to 25 GHz.
The :FFT:CENTer command sets the center frequency when FFT (Fast Fourier
Transform) is selected.
Query Syntax
:FFT:CENTer?
The :FFT:CENTer? query returns the current center frequency in Hertz.
Return Format
<frequency><NL>
<frequency> ::= the current center frequency in NR3 format. The range
of legal values is from -25 GHz to 25 GHz.
See Also • ":FFT:DISPlay" on page 262
• ":FFT:OFFSet" on page 263
• ":FFT:RANGe" on page 264
• ":FFT:REFerence" on page 265
• ":FFT:SCALe" on page 266
• ":FFT:SOURce1" on page 267
• ":FFT:SPAN" on page 268
• ":FFT:VTYPe" on page 269
• ":FFT:WINDow" on page 270
NOTE
After a *RST (Reset) or :AUToscale command, the values returned by the :FFT:CENTer? and
:FFT:SPAN? queries depend on the current :TIMebase:RANGe value. Once you change either
the :FFT:CENTer or :FFT:SPAN value, they no longer track the :TIMebase:RANGe value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals