EasyManuals Logo

Keysight Technologies InfiniiVision DSOX1204A User Manual

Keysight Technologies InfiniiVision DSOX1204A
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #590 background imageLoading...
Page #590 background image
590 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
27 :TRIGger Commands
:TRIGger:HOLDoff
(see page 798)
Command Syntax
:TRIGger:HOLDoff <holdoff_time>
<holdoff_time> ::= 60 ns to 10 s in NR3 format
The :TRIGger:HOLDoff command defines the holdoff time value in seconds.
Holdoff keeps a trigger from occurring until after a certain amount of time has
passed since the last trigger. This feature is valuable when a waveform crosses the
trigger level multiple times during one period of the waveform. Without holdoff,
the oscilloscope could trigger on each of the crossings, producing a confusing
waveform. With holdoff set correctly, the oscilloscope always triggers on the same
crossing. The correct holdoff setting is typically slightly less than one period.
Query Syntax
:TRIGger:HOLDoff?
The :TRIGger:HOLDoff? query returns the holdoff time value for the current trigger
mode.
Return Format
<holdoff_time><NL>
<holdoff_time> ::= the holdoff time value in seconds in NR3 format.
See Also • "Introduction to :TRIGger Commands" on page 585

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals