394 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:VTOP
(see page 798)
Command Syntax
:MEASure:VTOP [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VTOP command installs a screen measurement and starts a
waveform top value measurement.
Query Syntax
:MEASure:VTOP? [<source>]
The :MEASure:VTOP? query returns the vertical value at the top of the waveform.
The top value of the pulse is normally not the same as the maximum value.
Return Format
<value><NL>
<value> ::= vertical value at the top of the waveform in NR3 format
See Also • "Introduction to :MEASure Commands" on page 350
• ":MEASure:SOURce" on page 380
• ":MEASure:VMAX" on page 389
• ":MEASure:VAMPlitude" on page 386
• ":MEASure:VBASe" on page 388
NOTE
This query is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals