24 :SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth
(see page 798)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command sets the number of
8-bit bytes in the CAN data string. The number of bytes in the string can be
anywhere from 1 bytes to 8 bytes (64 bits). The value for these bytes is set by the
:SBUS<n>:CAN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth? query returns the current
CAN data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors • "-241, Hardware missing" on page 759
See Also • "Introduction to :TRIGger Commands" on page 585
• ":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 479
• ":SBUS<n>:CAN:SOURce" on page 476

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals