268 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
15 :FFT Commands
:FFT:SPAN
(see page 798)
Command Syntax
:FFT:SPAN <span>
<span> ::= the current frequency span in NR3 format. Legal values are
1 Hz to 100 GHz.
If you set the frequency span to a value outside of the legal range, the
step size is automatically set to the nearest legal value.
The :FFT:SPAN command sets the frequency span of the display (left graticule to
right graticule) when FFT (Fast Fourier Transform) is selected.
Query Syntax
:FFT:SPAN?
The :FFT:SPAN? query returns the current frequency span in Hertz.
Return Format
<span><NL>
<span> ::= the current frequency span in NR3 format. Legal values are 1
Hz to 100 GHz.
See Also • ":FFT:CENTer" on page 261
• ":FFT:DISPlay" on page 262
• ":FFT:OFFSet" on page 263
• ":FFT:RANGe" on page 264
• ":FFT:REFerence" on page 265
• ":FFT:SCALe" on page 266
• ":FFT:SOURce1" on page 267
• ":FFT:VTYPe" on page 269
• ":FFT:WINDow" on page 270
NOTE
After a *RST (Reset) or :AUToscale command, the values returned by the :FFT:CENTer? and
:FFT:SPAN? queries depend on the current :TIMebase:RANGe value. Once you change either
the :FFT:CENTer or :FFT:SPAN value, they no longer track the :TIMebase:RANGe value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals