:WAVeform Commands 28
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 665
:WAVeform:YINCrement
(see page 798)
Query Syntax
:WAVeform:YINCrement?
The :WAVeform:YINCrement? query returns the y-increment value in volts for the
currently specified source. This value is the voltage difference between
consecutive data values.
Return Format
<value><NL>
<value> ::= y-increment value in the current preamble in 32-bit
floating point NR3 format
See Also • "Introduction to :WAVeform Commands" on page 635
• ":WAVeform:PREamble" on page 649
Example Code • "Example Code" on page 650