482 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE
(see page 798)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE <value>
<value> ::= {STANdard | EXTended}
The :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE command sets the CAN identifier
mode. STANdard selects the standard 11-bit identifier. EXTended selects the
extended 29-bit identifier. The CAN identifier is set by the
:SBUS<n>:CAN:TRIGger:PATTern:ID command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE?
The :SBUS<n>:CAN:TRIGger:PATTern:ID:MODE? query returns the current setting
of the CAN identifier mode.
Return Format
<value><NL>
<value> ::= {STAN | EXT}
Errors • "-241, Hardware missing" on page 759
See Also • "Introduction to :TRIGger Commands" on page 585
• ":SBUS<n>:MODE" on page 465
• ":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 479
• ":SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth" on page 480
• ":SBUS<n>:CAN:TRIGger:PATTern:ID" on page 481

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals