:EXTernal Trigger Commands 14
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 255
:EXTernal:POSition
(see page 798)
Command Syntax
:EXTernal:POSition <value>
<value> ::= Zero in NR3 format.
In the InfiniiVision DSOX1204A/G oscilloscopes, the :EXTernal:POSition command
can be set to zero only (since the Ext Trig input cannot be displayed as a digital
waveform).
Query Syntax
:EXTernal:POSition?
The :EXTernal:POSition? query returns zero.
Return Format
<value><NL>
See Also • ":EXTernal:DISPlay" on page 252
• ":EXTernal:LEVel" on page 254