644 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
28 :WAVeform Commands
:WAVeform:FORMat
(see page 798)
Command Syntax
:WAVeform:FORMat <value>
<value> ::= {WORD | BYTE | ASCii}
The :WAVeform:FORMat command sets the data transmission mode for waveform
data points. This command controls how the data is formatted when sent from the
oscilloscope.
• ASCii formatted data converts the internal integer data values to real Y-axis
values. Values are transferred as ASCii digits in floating point notation,
separated by commas.
ASCII formatted data is transferred ASCii text.
• WORD formatted data transfers 16-bit data as two bytes. The
:WAVeform:BYTeorder command can be used to specify whether the upper or
lower byte is transmitted first. The default (no command sent) is that the upper
byte transmitted first.
• BYTE formatted data is transferred as 8-bit bytes.
Query Syntax
:WAVeform:FORMat?
The :WAVeform:FORMat query returns the current output format for the transfer of
waveform data.
Return Format
<value><NL>
<value> ::= {WORD | BYTE | ASC}
See Also • "Introduction to :WAVeform Commands" on page 635
• ":WAVeform:BYTeorder" on page 640
• ":WAVeform:SOURce" on page 654
• ":WAVeform:DATA" on page 642
• ":WAVeform:PREamble" on page 649
Example Code • "Example Code" on page 654

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals