698 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
29 :WGEN Commands
:WGEN:VOLTage:LOW
(see page 798)
Command Syntax
:WGEN:VOLTage:LOW <low>
<low> ::= low-level voltage in volts, in NR3 format
For all waveforms except DC, the :WGEN:VOLTage:LOW command specifies the
waveform's low-level voltage. Use the :WGEN:VOLTage:HIGH command to specify
the high-level voltage.
You can also specify the high-level and low-level voltages indirectly using the
:WGEN:VOLTage and :WGEN:VOLTage:OFFSet commands. For example, a
high-level voltage of 4 V and a low-level voltage of -1 V is the same as an
amplitude of 5 V and an offset of 1 V.
Query Syntax
:WGEN:VOLTage:LOW?
The :WGEN:VOLTage:LOW? query returns the currently specified waveform
low-level voltage.
Return Format
<low><NL>
<low> ::= low-level voltage in volts, in NR3 format
See Also • "Introduction to :WGEN Commands" on page 671
• ":WGEN:FUNCtion" on page 673
• ":WGEN:VOLTage:LOW" on page 698
• ":WGEN:VOLTage" on page 696
• ":WGEN:VOLTage:OFFSet" on page 699