:WMEMory<r> Commands 30
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 709
:WMEMory<r>:YRANge
(see page 798)
Command Syntax
:WMEMory<r>:YRANge <range>[<suffix>]
<r> ::= 1-2 in NR1 format
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
The :WMEMory<r>:YRANge command defines the full-scale vertical axis of the
selected reference waveform.
Legal values for the range are copied from the original source waveform (that is,
the analog channel or math function waveform that was originally saved as a
reference waveform).
Query Syntax
:WMEMory<r>:YRANge?
The :WMEMory<r>:YRANge? query returns the current full-scale range setting for
the specified reference waveform.
Return Format
<range><NL>
<range> ::= vertical full-scale range value in NR3 format
See Also Chapter 30, “:WMEMory<r> Commands,” starting on page 701
":WMEMory<r>:DISPlay" on page 704
":WMEMory<r>:YOFFset" on page 708
":WMEMory<r>:SKEW" on page 707
":WMEMory<r>:YSCale" on page 710

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals