Obsolete and Discontinued Commands 31
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 735
:MEASure:TMAX
(see page 798)
Command Syntax
:MEASure:TMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMAX command installs a screen measurement and starts an
X-at-Max-Y measurement on the selected waveform. If the optional source is
specified, the current source is modified.
Query Syntax
:MEASure:TMAX? [<source>]
The :MEASure:TMAX? query returns the horizontal axis value at which the
maximum vertical value occurs on the current source. If the optional source is
specified, the current source is modified. If all channels are off, the query returns
9.9E+37.
Return Format
<value><NL>
<value> ::= time at maximum in NR3 format
See Also • "Introduction to :MEASure Commands" on page 350
• ":MEASure:TMIN" on page 736
• ":MEASure:XMAX" on page 396
• ":MEASure:XMIN" on page 397
The :MEASure:TMAX command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:XMAX command (see page 396) instead.