Obsolete and Discontinued Commands 31
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 755
:TIMebase:DELay
(see page 798)
Command Syntax
:TIMebase:DELay <delay_value>
<delay_value> ::= time in seconds from trigger to the delay reference
point on the screen.
The valid range for delay settings depends on the time/division
setting for the main time base.
The :TIMebase:DELay command sets the main time base delay. This delay is the
time between the trigger event and the delay reference point on the screen. The
delay reference point is set with the :TIMebase:REFerence command (see
page 578).
Query Syntax
:TIMebase:DELay?
The :TIMebase:DELay query returns the current delay value.
Return Format
<delay_value><NL>
<delay_value> ::= time from trigger to display reference in seconds
in NR3 format.
Example Code ' TIMEBASE_DELAY - Sets the time base delay. This delay
' is the internal time between the trigger event and the
' onscreen delay reference point.
' Set time base delay to 0.0.
myScope.WriteString ":TIMEBASE:DELAY 0.0"
See complete example programs at: Chapter 36, “Programming Examples,”
starting on page 807
NOTE
The :TIMebase:DELay command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :TIMebase:POSition command (see page 576) instead.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals