278 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
16 :FRANalysis Commands
:FRANalysis:FREQuency:SINGle
(see page 798)
Command Syntax
:FRANalysis:FREQuency:SINGle <value>[suffix]
<value> ::= between 10 Hz and 20 MHz with a resolution of 0.1 Hz
in NR3 format
[suffix] ::= {Hz | kHz| MHz}
The :FRANalysis:FREQuency:SINGle command command sets the single frequency
value.
Query Syntax
:FRANalysis:FREQuency:SINGle?
The :FRANalysis:FREQuency:SINGle? query returns the single frequency setting.
Return Format
<value><NL>
See Also • ":FRANalysis:DATA" on page 275
• ":FRANalysis:ENABle" on page 276
• ":FRANalysis:SWEep:POINts" on page 284
• ":FRANalysis:FREQuency:MODE" on page 277
• ":FRANalysis:RUN" on page 281
• ":FRANalysis:SOURce:INPut" on page 282
• ":FRANalysis:SOURce:OUTPut" on page 283
• ":FRANalysis:WGEN:LOAD" on page 286
• ":FRANalysis:WGEN:VOLTage" on page 287
The SINGle mode is available through the remote programming interface only and is not
available when using the front panel graphical user interface (GUI).