:CHANnel<n> Commands 10
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 205
:CHANnel<n>:BWLimit
(see page 798)
Command Syntax
:CHANnel<n>:BWLimit <bwlimit>
<bwlimit> ::= {{1 | ON} | {0 | OFF}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:BWLimit command controls an internal low-pass filter. When
the filter is on, the bandwidth of the specified channel is limited to approximately
25 MHz.
Query Syntax
:CHANnel<n>:BWLimit?
The :CHANnel<n>:BWLimit? query returns the current setting of the low-pass
filter.
Return Format
<bwlimit><NL>
<bwlimit> ::= {1 | 0}
See Also • "Introduction to :CHANnel<n> Commands" on page 203