EasyManuals Logo

Keysight Technologies InfiniiVision DSOX1204A User Manual

Keysight Technologies InfiniiVision DSOX1204A
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #609 background imageLoading...
Page #609 background image
:TRIGger Commands 27
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 609
:TRIGger:GLITch:RANGe
(see page 798)
Command Syntax
:TRIGger:GLITch:RANGe <less_than_time>[suffix],
<greater_than_time>[suffix]
<less_than_time> ::= (15 ns - 10 seconds) in NR3 format
<greater_than_time> ::= (10 ns - 9.99 seconds) in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:RANGe command sets the pulse width duration for the
selected :TRIGger:GLITch:SOURce. You can enter the parameters in any order —
the smaller value becomes the <greater_than_time> and the larger value becomes
the <less_than_time>.
Query Syntax
:TRIGger:GLITch:RANGe?
The :TRIGger:GLITch:RANGe? query returns the pulse width duration time for
:TRIGger:GLITch:SOURce.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also • "Introduction to :TRIGger Commands" on page 585
• ":TRIGger:GLITch:SOURce" on page 610
• ":TRIGger:GLITch:QUALifier" on page 608
• ":TRIGger:MODE" on page 594

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals