:MARKer Commands 19
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 339
:MARKer:YUNits
(see page 798)
Command Syntax
:MARKer:YUNits <units>
<units> ::= {BASE | PERCent}
The :MARKer:YUNits command sets the Y cursors units:
• BASE — for making measurements in the units associated with the cursors
source.
• PERCent — for making ratio measurements. Use the :MARKer:YUNits:USE
command to set the current Y1 location as 0 percent and the current Y2
location as 100 percent.
Changing Y units affects the input and output values of the :MARKer:Y1Position,
:MARKer:Y2Position, and :MARKer:YDELta commands/queries.
Query Syntax
:MARKer:YUNits?
The :MARKer:YUNits? query returns the current Y cursors units.
Return Format
<units><NL>
<units> ::= {BASE | PERC}
See Also • "Introduction to :MARKer Commands" on page 326
• ":MARKer:YUNits:USE" on page 340
• ":MARKer:X1Y1source" on page 330
• ":MARKer:X2Y2source" on page 332
• ":MEASure:SOURce" on page 380
• ":MARKer:Y1Position" on page 336
• ":MARKer:Y2Position" on page 337
• ":MARKer:YDELta" on page 338

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals