Page #164 background image
164 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
6 Root (:) Commands
:TER (Trigger Event Register)
(see page 798)
Query Syntax
:TER?
The :TER? query reads the Trigger Event Register. After the Trigger Event Register
is read, it is cleared. A one indicates a trigger has occurred. A zero indicates a
trigger has not occurred.
The Trigger Event Register is summarized in the TRG bit of the Status Byte
Register (STB). A Service Request (SRQ) can be generated when the TRG bit of the
Status Byte transitions, and the TRG bit is set in the Service Request Enable
register. The Trigger Event Register must be cleared each time you want a new
service request to be generated.
Return Format
<value><NL>
<value> ::= {1 | 0}; a 16-bit integer in NR1 format.
See Also • "Introduction to Root (:) Commands" on page 135
• "*SRE (Service Request Enable)" on page 125
• "*STB (Read Status Byte)" on page 127

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals