31 Obsolete and Discontinued Commands
:FUNCtion:SOURce
(see page 798)
Command Syntax
:FUNCtion:SOURce <value>
<value> ::= {CHANnel<n> | ADD | SUBTract | MULTiply}
<n> ::= 1 to (# analog channels) in NR1 format
The :FUNCtion:SOURce command is only used when an FFT (Fast Fourier
Transform) operation is selected (see the:FUNCtion:OPERation command for more
information about selecting an operation). The :FUNCtion:SOURce command
selects the source for function operations. Choose CHANnel<n>, or ADD, SUBT, or
MULT to specify the desired source for FFT operations specified by the
:FUNCtion:OPERation command.
Query Syntax
:FUNCtion:SOURce?
The :FUNCtion:SOURce? query returns the current source for function operations.
Return Format
<value><NL>
<value> ::= {CHAN<n> | ADD | SUBT | MULT}
<n> ::= 1 to (# analog channels) in NR1 format
See Also • "Introduction to :FUNCtion Commands" on page 291
• ":FUNCtion:OPERation" on page 303
The :FUNCtion:SOURce command is an obsolete command provided for compatibility to