EasyManuals Logo

Keysight Technologies InfiniiVision DSOX1204A User Manual

Keysight Technologies InfiniiVision DSOX1204A
946 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #497 background imageLoading...
Page #497 background image
:SBUS<n> Commands 24
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 497
:SBUS<n>:LIN:SIGNal:BAUDrate
(see page 798)
Command Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
The :SBUS<n>:LIN:SIGNal:BAUDrate command sets the standard baud rate of the
LIN signal from 2400 b/s to 625 kb/s in 100 b/s increments. If you enter a baud
rate that is not divisible by 100 b/s, the baud rate is set to the nearest baud rate
divisible by 100 b/s.
Query Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate?
The :SBUS<n>:LIN:SIGNal:BAUDrate? query returns the current LIN baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
See Also • "Introduction to :TRIGger Commands" on page 585
• ":TRIGger:MODE" on page 594
• ":SBUS<n>:LIN:TRIGger" on page 501
• ":SBUS<n>:LIN:SOURce" on page 498

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals