218 Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide
10 :CHANnel<n> Commands
:CHANnel<n>:RANGe
(see page 798)
Command Syntax
:CHANnel<n>:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:RANGe command defines the full-scale vertical axis of the
selected channel. When using 1:1 probe attenuation, legal values for the range are
from 8 mV to 40 V.
If the probe attenuation is changed, the range value is multiplied by the probe
attenuation factor.
Query Syntax
:CHANnel<n>:RANGe?
The :CHANnel<n>:RANGe? query returns the current full-scale range setting for
the specified channel.
Return Format
<range_argument><NL>
<range_argument> ::= vertical full-scale range value in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 203
":CHANnel<n>:SCALe" on page 219
":CHANnel<n>:PROBe" on page 212
Example Code
' CHANNEL_RANGE - Sets the full scale vertical range in volts. The
' range value is 8 times the volts per division.
myScope.WriteString ":CHANnel1:RANGe 8" ' Set the vertical range to
8 volts.
See complete example programs at: Chapter 36, “Programming Examples,”
starting on page 807

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals