:FUNCtion Commands 17
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 301
:FUNCtion:GOFT:SOURce2
(see page 798)
Command Syntax
:FUNCtion:GOFT:SOURce2 <value>
<value> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :FUNCtion:GOFT:SOURce2 command selects the second input channel for the
g(t) source that can be used as the input to the FFT function.
Query Syntax
:FUNCtion:GOFT:SOURce2?
The :FUNCtion:GOFT:SOURce2? query returns the current selection for the second
input channel for the g(t) source.
Return Format
<value><NL>
<value> ::= CHAN<n>
<n> ::= 1 to (# analog channels) in NR1 format
See Also • "Introduction to :FUNCtion Commands" on page 291
• ":FUNCtion:GOFT:SOURce1" on page 300
• ":FUNCtion:GOFT:OPERation" on page 299