:MEASure Commands 20
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 347
n/a :MEASure:TVALue?
<value>,
[<slope>]<occurrence>
[,<source>] (see
page 384)
<value> ::= voltage level that
the waveform must cross.
<slope> ::= direction of the
waveform when <value> is crossed.
<occurrence> ::= transitions
reported.
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= time in
seconds of specified voltage
crossing in NR3 format
:MEASure:VAMPlitude
[<source>] (see
page 386)
:MEASure:VAMPlitude?
[<source>] (see
page 386)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= the amplitude
of the selected waveform in volts
in NR3 format
:MEASure:VAVerage
[<interval>][,][<sour
ce>] (see page 387)
:MEASure:VAVerage?
[<interval>][,][<sour
ce>] (see page 387)
<interval> ::= {CYCLe | DISPlay}
<source> ::= {CHANnel<n> |
FUNCtion | MATH | FFT |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= calculated
average voltage in NR3 format
:MEASure:VBASe
[<source>] (see
page 388)
:MEASure:VBASe?
[<source>] (see
page 388)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<base_voltage> ::= voltage at the
base of the selected waveform in
NR3 format
Table 65 :MEASure Commands Summary (continued)
Command Query Options and Query Returns