:MEASure Commands 20
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 349
n/a :MEASure:VTIMe?
<vtime>[,<source>]
(see page 393)
<vtime> ::= displayed time from
trigger in seconds in NR3 format
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r> |
EXTernal}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= voltage at the
specified time in NR3 format
:MEASure:VTOP
[<source>] (see
page 394)
:MEASure:VTOP?
[<source>] (see
page 394)
<source> ::= {CHANnel<n> |
FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1-2 in NR1 format
<return_value> ::= voltage at the
top of the waveform in NR3 format
:MEASure:WINDow
<type> (see page 395)
:MEASure:WINDow? (see
page 395)
<type> ::= {MAIN | ZOOM | AUTO}
:MEASure:XMAX
[<source>] (see
page 396)
:MEASure:XMAX?
[<source>] (see
page 396)
<source> ::= {CHANnel<n> |
FUNCtion | FFT | MATH |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= horizontal
value of the maximum in NR3
format
:MEASure:XMIN
[<source>] (see
page 397)
:MEASure:XMIN?
[<source>] (see
page 397)
<source> ::= {CHANnel<n> |
FUNCtion | FFT | MATH |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= horizontal
value of the maximum in NR3
format
Table 65 :MEASure Commands Summary (continued)
Command Query Options and Query Returns