:MEASure Commands 20
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 357
<occurrence> ::= integer
This command defines the behavior of the :MEASure:DELay? query by specifying
the start and stop edge to be used. <edge_spec1> specifies the slope and edge
number on source1. <edge_spec2> specifies the slope and edge number on
source2. The measurement is taken as:
delay = t(<edge_spec2>) - t(<edge_spec1>)
:MEASure:DEFine
THResholds
Command Syntax
:MEASure:DEFine THResholds,<threshold spec>[,<source>]
<threshold spec> ::= {STANdard}
| {<threshold mode>,<upper>,<middle>,<lower>}
<threshold mode> ::= {PERCent | ABSolute}
for <threshold mode> = PERCent:
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
and lower threshold percentage values
between Vbase and Vtop in NR3 format.
for <threshold mode> = ABSolute:
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
and lower threshold absolute values in
NR3 format.
• STANdard threshold specification sets the lower, middle, and upper
measurement thresholds to 10%, 50%, and 90% values between Vbase and
Vtop.
• Threshold mode PERCent sets the measurement thresholds to any user-defined
percentages between 5% and 95% of values between Vbase and Vtop.
• Threshold mode ABSolute sets the measurement thresholds to absolute values.
ABSolute thresholds are dependent on channel scaling (:CHANnel<n>:RANGe
or ":CHANnel<n>:SCALe" on page 219:CHANnel<n>:SCALe), probe attenuation
(:CHANnel<n>:PROBe), and probe units (:CHANnel<n>:UNITs). Always set
these values first before setting ABSolute thresholds.
Query Syntax
:MEASure:DEFine? <meas_spec>[,<source>]
<meas_spec> ::= {DELay | THResholds}
The :MEASure:DEFine? query returns the current edge specification for the delay
measurements setup or the current specification for the thresholds setup.
Return Format for <meas_spec> = DELay:
NOTE
The :MEASure:DELay command and the front-panel delay measurement use an auto-edge
selection method to determine the actual edge used for the measurement. The
:MEASure:DEFine command has no effect on these delay measurements. The edges specified
by the :MEASure:DEFine command only define the edges used by the :MEASure:DELay? query.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision DSOX1204A and is the answer not in the manual?

Keysight Technologies InfiniiVision DSOX1204A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision DSOX1204A
CategoryTest Equipment
LanguageEnglish

Related product manuals