Status Reporting 33
Keysight InfiniiVision DSOX1204A/G Oscilloscopes Programmer's Guide 769
To generate a service request (SRQ) interrupt to an external controller, at least one
bit in the Status Byte Register must be enabled. These bits are enabled by using
the *SRE common command to set the corresponding bit in the Service Request
Enable Register. These enabled bits can then set RQS and MSS (bit 6) in the
Status Byte Register.