No. 99MBG518A
G-47
PART G : 4 Results
No. Name Details
Print button Outputs the displayed data to an external device.
Set the external output settings to Manual or Automatic to output data to
an external device.
For details about the external output settings, see "3.3 External Out-
put" (page G-30).
■When the group settings are enabled
No. Name Details
Change data buttons Touch this button to switch the display to show the data for the previ-
ous/next test results.
Group Displays the group number on the left and the number of tested
measuring points/total number of measuring points on the right.
Touch this button to change the group number of the test results to
display.
Conversion Displays the conversion results and conversion scale for the test
results if the conversion setting is set to a value other than OFF.
The conversion value is displayed on the left and the conversion
scale on the right.
GO/NG judgment Displays GO/+NG/-NG for the test results if judgment is enabled.
GO: Passed
+NG: Upper limit failed
-NG: Lower limit failed
Hardness value
(average)
Displays the average value for the number of measuring points of
the specimen that is displayed.
Displays the average value excluding the maximum and minimum
values if the group settings are set to exclude the maximum and
minimum values.
Scale Displays the hardness scale during a test.
Test force Displays the total test force during the test.
Correction icon Displays an icon if user correction or curve correction is enabled.
Test number list Displays a list of the test numbers.
Change test number Changes the display to the previous/next result in the test number
list if all of the test results are not displayed.